Single channel operation
Precautions while using Cathode Ray Oscilloscope
Single channel operation
1. Set the coupling of channel 1 input to DC and connect the signal generator to this
input. Note that the outer connectors of the coaxial sockets are connected to
earth. This means that if one of the signal generators terminals is also grounded,
the two grounded contacts must be the ones which are connected together. .
2. Set the generator to produce a sine wave of frequency 1 kHz and amplitude 1 V.
Some signal generators specify the peak to peak amplitude or alternatively the
r.m.s voltage. In these cases use 2 V p p or 1.41 V respectively .
3. If the signal generator has the option of adding a DC offset to the signal set this to
0V so that the signal is purely AC.
4. Reset the oscilloscope time base to 1 ms per division and set the channel gain to
0.5 V per division.
Ensuring the cro is calibrated correctly
Heat light and sound
Delayed sweep
Cathode ray tube experiment
Time base
Cro operation
Cathode ray tube
Measuring the input signal
Trigger
Precautions
Lichtenberg figures
Checking that the cro is triggered correctly