Balance the tubes in the rotor
Precautions while using Centrifuge
Balance the tubes in the rotor
If you want to run a tube with 10 mL of liquid, put another tube with 10 mL of water in the opposing hole on the rotor (see photo, below). If the liquid has a higher or lower density than water, you must balance the tubes by mass, not volume. The total mass of each tube should be as close as possible this becomes increasingly important at very high rotor speeds. Running a centrifuge with unbalanced load could permanently damage the centrifuge. It could also cause injury to you or someone else.
Know your rotors and maintain a log book
Glass centrifuge tubes
A ThermoFisher laboratory bench top centrifuge
If you see it wobbling or shaking pull the plug
After centrifugation
Lids
General Before centrifugation
The work surface must be level and firm
Safe Procedures for Centrifugation
Low speed centrifuges
Wear a face shield and or safety goggles
Do not bump jar or move the centrifuge while the rotor is spinning More ...